| Capacity |
: |
224 litres
|
| Internal Dimensions |
: |
600 x 535 x 700mm (WxDxH)
|
| External Dimensions |
:
|
850 x 1460 x 1516mm (WxDxH)) - for height with castors add 47mm
|
| Temperature Range |
:
|
-40°C to +180°C
|
| Fluctuation |
:
|
±0.25șC
±0.3șC
|
| Temperature Variation Rate (according to IEC60068-3-5) |
:
|
5șC/min from -40șC to +180șC 3.5șC/min from +180șC to -40șC (average rates with no load and ambient temperature of +20șC to +22șC)
|
| Relative Humidity Range |
:
|
From 10% to 98% in the range +5șC to +95șC
|
| Dewpoint Range |
:
|
+2șC to +94șC for continuous tests, -20șC to +2șC for limited period based on starting conditions and foreseen testing time.
|
| Relative Humidity Stability |
:
|
±1%
±3% and in all cases not less than ±0.25șC within the psychrometric table
|
| Control |
:
|
By ”PLC and colour KeyKratos touchscreen placed on the door (24 touch areas, 1024 dynamic tests, IP65 frontal panel protection grade). It allows programming of the T and RH cycles and control of all chamber functions and alarms.
|
| Dissapation at 025șC |
:
|
400 W
|
| Electrical Power |
:
|
7.7kW (4.8kW average)
|
| Voltage |
:
|
400V +6%/-10%/50Hz/3ph + N + G
|
| Noise |
:
|
59 db(A) at 1m from the front of the chamber in a non-reverberating room
|
| Weight |
:
|
490 kg
|
| Included Accessories |
:
|
450 x 450mm Multipane inspection window
Castors assembly
Self-feeding system (mains water through chamber embedded softener)
1 lateral porthole (dia. 80mm)
1 Internal shelf
Internal lamp
Air condenser
RS232 interface (or RS485 on request)
Digital undertemperature/overtemperature switch with independent probe
Connection to UPS
False ceiling to prevent water dripping on the test object
|
| The software source code and the relative documentation are not included.
For futher information please contact Sharon |